Recall Z-0919-2007 ยท LEH Radioimmunoassay, Vancomycin

Roche Diagnostics Corp.

Roche VANC2, Online TDM Vancomycin, for use with the Roche/Hitachi Cobas c 501analyzer; Catalog No.04491050190.

DateJun 9, 2007 (initiated Apr 30, 2007)
StatusTerminated
ReasonProbe Wash: Reported results may have a positive bias because inadequate analyzer probe wash instructions may result in carry-over from the following assays: Creatinine Kinase L and Creatine Kinase-MBL.
Root causeOther
ActionConsignees were issued a recall letter dated 4/30/07 instructing them to perform a R2 reagent probe wash when using CKL and CKMBL with the vancomycin or valproic acid assays.
Quantity662
DistributionNationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060586ENZYME IMMUNOASSAY, VANCOMYCIN
Roche Diagnostics Corp., May 11, 2006