Recall Z-0917-2007 ยท KLB Radioimmunoassay, Tobramycin

Roche Diagnostics Corp.

Roche Online TDM Tobramycin, for use on Roche/Hitachi analyzers 911,912, 917 and Modular P; Catalog No. 04642511190. (Note: Not distributed within the United States)

DateJun 9, 2007 (initiated Apr 5, 2007)
StatusTerminated
ReasonReported results may have a positive bias of up to 113% because inadequate cleaning instructions may result in carry-over of the following reagents: Online TDM carbamazepine, Online TDM gentamycin, Online TDM phenytoin, Online DAT cocaine II, Online DAT methadone II, Online DAT opiates II and Online DAT THC II.
Root causeOther
ActionForeign affiliates were notified via email letter dated 4/5/07 requesting that to notify their customers of the need to implement a special wash cycle to prevent reagent carry over.
Quantity474 total (Z-0917-0918-2007)
DistributionInternational Distribution ONLY-Australia, Canada and Germany.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060853ONLINE TDM TOBRAMYCIN
Roche Diagnostics Corp., Jun 9, 2006