Recall Z-0887-2007 ยท MJK Dna Probe, Trichomonas Vaginalis

Medtronic Emergency Response Systems, Inc.

Batteries used with the LIFEPAK 20 defibrillator/monitor

DateJul 4, 2007 (initiated Mar 30, 2007)
StatusTerminated
ReasonDefibrillator internal battery cell may contain a short which could reduce battery capacity and cause the device to shut down without warning while operating on backup (DC) battery power.
Root causeOther
ActionFirm sent letters on 03/30/2007 dated 03/2007 via certified return receipt to consignees with affected LIFEPAK 20 devices. The firm advised consignees to use the Operator's Checklist in the Operating Instructions to perform daily user tests and function checks while operating the defibrillator on battery power. In addition consignees are advised to replace their battery every 1 to 2 years.
Quantity3002
DistributionDistribution is worldwide. The device is designed to be used in healthcare provider facilities such as hospitals and clinic settings.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.