ClearanceBrief
Recall Z-0790-2007 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Abbott Laboratories, Inc

ARCHITECT i2000 Processing Module, for in vitro diagnostics, List Number 08C89-01; manufactured by Abbot Laboratories, Irving, TX.

DateMay 9, 2007 (initiated Mar 28, 2007)
StatusTerminated
ReasonIf the ARCHITECT i2000/i2000sr System becomes contaminated with microbes generating folate-like by-products, the ARCHITECT Folate assay may experience calibration failures and/or shifts in results.
Root causeOther
ActionNotification Letters were sent to all customers who received the ARCHITECT Folate Reagent on 03/28/07. Customers were provided with instructions on how to conduct a Wash Buffer Unload procedure and were requested to perform this action daily. Customers were also asked to contact their local Abbott Customer Support Rep to schedule an Internal Decontamination procedure on the ARCHITECT System. Firm recalling to the user (laboratory) level.
Quantityn/a
DistributionFolate Reagent distributed to the following countries: Australia, Austria, Belgium, Bosnia and Herzegovina, Brazil, Canada, Chile, China, Colombia, Croatia, Czech Republic, Denmark, Dominican Republic, Finland, France, Greece, Hungary, Ireland, Israel, Italy, Jordan, Lebanon, Luxembourg, Netherlands
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K983212ABBOTT ARCHITECT TESTOSTERONE
Abbott Laboratories, Nov 18, 1998