Recall Z-1048-2007 ยท NIK Defibrillator, Automatic Implantable Cardioverter, With Cardiac Resynchronization (Crt-D)

Boston Scientific CRM Corp

These models are not available in the US. CONTAK RENEWAL 4 (model H190, H195). CONTAK RENEWAL 4 HE (model H197, H199) Cardiac Resynchronization Therapy Defibrillator (CRT-D). Guidant Corporation, 4100 Hamline Avenue North, St. Paul, MN 55112-5798 USA.

DateJul 11, 2007 (initiated Mar 23, 2007)
StatusTerminated
ReasonBoston Scientific has recently identified low-voltage capacitors from a former supplier that may be subject to degradation at a low rate of occurrence. These capacitors may cause accelerated battery depletion and may reduce the time between elective replacement indicator (ERI) and end of life (EOL) to less than 3 months.
Root causeOther
ActionAn Urgent Medical Devivce Information letter dated April 05, 2007 was sent to implanting and following physicians. The letter describes the issue, identifies patients with affected devices and provides recommendations.
Quantity6,700
DistributionWorldwide, including USA, Argentina, Aruba, Australia, Austria, Barbados, Belgium, Bermuda, Brazil, Canada, Chile, Costa Rica, Cyprus, Czech Republic, Czechoslovakia, Denmark, Dominican Republic, Ecuador, Egypt, Finland, France, Germany, Greece, Guam, Hong Kong, Hungary, Iceland, India, Iran, Irelan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.