Recall Z-0739-2007 ยท IZW Collimator, Automatic, Radiographic

Philips Medical Systems North America Co. Phillips

Philips Collimators, Model numbers: 9896 010 00614, 9896 010 00615, 9804 607 00301, and 9804 602 83501. Product is a Collimator/ Beam Limiting Device and are replacement components for the Bucky Diagnost TH X-ray system.

DateJun 14, 2007 (initiated Mar 15, 2007)
StatusTerminated
ReasonThe Collimator field-defining light beam does not meet the light field contrast ratio requirement of 4:1. System is non-compliant with performance standard (1020.31 (d)(2)(iii) 21 CFR
Root causeOther
ActionFirm issued letter on 03/20/2007 to consignees advising them that firm representatives will contact consignee to replace collimators at no charge.
Quantity6
DistributionNationwide; Units were distributed to hospitals, radiologists, and health care clinics in IA, LA, ND, PA, and TX, and Washington DC.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.