ClearanceBrief
Recall Z-0853-2007 ยท GGP Test, Qualitative And Quantitative Factor Deficiency

Diagnostica Stago, Inc.

STA Liatest VWF:Ag, Product Catalog Number : 0518. This is a Immuno-Turbidimetric Assay of von Willebrand Factor.

DateJul 3, 2007 (initiated Mar 13, 2007)
StatusTerminated
ReasonA dose-hook effect is observed starting from VWF:AG levels of 300%, while the package insert indicates that no dose-hook effect has been observed with VWF:Ag levels up to 800%.
Root causeOther
ActionCustomers were notified on March 13, 2007 by letter, first class mail.
Quantity1,098 units
DistributionHospitals and laboratories Nationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall