Recall Z-0751-2007 ยท IZL System, X-Ray, Mobile
Siemens Medical Solutions USA, Inc
Mobilett XP, mobile X-ray system, Model number 1818363
| Date | Apr 19, 2007 (initiated Mar 21, 2007) |
| Status | Terminated |
| Reason | Mounting bolts for tank fork assembly and collimator flange may become loose. |
| Root cause | Other |
| Action | Siemens Service Personnel are visiting each consignee, beginning March 21, 2007, to replace the required mounting bolts and tighten them according to required torque specifications. |
| Quantity | 5 units |
| Distribution | The products were shipped to medical facilities nationwide. |