Recall Z-0693-2007 ยท DZE Implant, Endosseous, Root-Form

BioHorizons Implant Systems Inc

BioHorizons 0.050 Hex Driver, Regular, REF 300-350, BioHorizons Implant Systems, Inc., Birmingham, AL 35243, non-sterile, Rx only

DateApr 13, 2007 (initiated Mar 12, 2007)
StatusTerminated
ReasonPremature Wear; The material from which the hex driver was made was improperly hardened during manufacturing and could lead to premature wear.
Root causeOther
ActionLetters and replacement drivers were sent to the consignees (both domestic and abroad) on 3/12/2007 explaining the problem and requesting that the consignee discard the instrument and use the replacement. They were delivered via United Parcel Service (UPS) 3rd day package delivery.
Quantity495 units
DistributionWorldwide Distribution -- USA and countries of Canada, Norway, Australia, Spain, Brazil, Pakistan, Israel, Latvia, Japan, Columbia, Thailand, Turkey, Italy, South Korea, India, Taiwan, Ukraine, Russia, United Arab Emirates, Saudi Arabia, Hong Kong, Romania, UK, Kuwait, and Lithuania
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K963951AS 123 PROSTHETIC INSTRUMENTATION SYSTEM
BioHorizons Implant Systems, Inc., Dec 31, 1996