Recall Z-1790-2012 ยท KPR System, X-Ray, Stationary

General Electric Med Systems LLC

GE Proteus XR/a Radiographic System. GE Medical Systems-Americas Milwaukee, WI. Model Number: 2259988. Product is indicated for use in generating radiographic images of human anatomy in all general-purpose diagnostic procedures. This device is not intended for mammographic applications.

DateJun 8, 2012 (initiated May 31, 2006)
StatusTerminated
ReasonAn incident was reported that an operator's finger was pinched between the fixed roller and the gliding tabletop of a Proteus XR/a Table while moving the tabletop with the thumb placed over its edge and fingers underneath
Root causeDevice Design
ActionGE Healthcare sent an "URGENT SAFETY INFORMATION" letter dated May 31, 2006 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter included new operator instructions and warning labels.
Quantity3,796 units
DistributionWorldwide Distribution.