Recall Z-1789-2012 ยท MQB Solid State X-Ray Imager (Flat Panel/Digital Imager)

General Electric Med Systems LLC

GE Healthcare Definium 8000 Digital Radiographic Systems : GE Healthcare, 3000 North Grandview , Waukesha, WI, 53188 Model Number: 5131070. The Revolution XR/d with Tomosynthesis is intended for use in generating Tomographic images of human anatomy. It is not intended for mammographic use.

DateJun 8, 2012 (initiated May 31, 2006)
StatusTerminated
ReasonAn incident was reported that an operator's finger was pinched between the fixed roller and the gliding tabletop of a Proteus XR/a Table while moving the tabletop with the thumb placed over its edge and fingers underneath
Root causeDevice Design
ActionGE Healthcare sent an "URGENT SAFETY INFORMATION" letter dated May 31, 2006 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter included new operator instructions and warning labels.
Quantity3,796 units
DistributionWorldwide Distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K012389REVOLUTION XR/D DIGITAL RADIOGRAPHIC IMAGING SYSTEM
Ge Medical Systems, Inc., Aug 10, 2001