Recall Z-0727-2007 ยท FSA Lift, Patient, Non-Ac-Powered

Joerns Healthcare Inc.

Sunrise Medical-Hoyer Advance Patient Lifts (Portable and Folding Patient Lift) Sunrise Medical, Inc.

DateApr 7, 2007 (initiated Feb 21, 2007)
StatusTerminated
ReasonUnapproved Design Change (by base supplier) consisting of a reduction in the number of mast-to-base welds (from four welds to two welds).
Root causeOther
ActionRecall initiated by letter dated 2/21/2007. Sunrise Medical will schedule onsite appointments to complete the base replacement.
Quantity581 units
DistributionWorldwide Distribution --- USA including states of AK, AL, AR, AZ, CA, CO, CT, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NH, NJ, NM, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV and country of Canada.