ClearanceBrief
Recall Z-0559-2007 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

GE OEC Medical Systems, Inc

OEC 9800MD Motorized C-arm System, Fluoroscopic X-ray System,

DateMar 22, 2007 (initiated Feb 9, 2007)
StatusTerminated
ReasonDuring routine service, a cover may have been installed without the required proper lead shielding.
Root causeOther
ActionConsignees were notified by telephone on 02/09/2007 and told to discontinue use of the units until they could be inspected by GE personnel. GE personnel visited all 8 sites on 02/12/2007.
Quantity11 units of all models
DistributionNationwide to AL, AZ, IA, OH, TX, VA, and WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K021049OEC 9800 PLUS DIGITAL MOBILE IMAGING SYSTEM
Ge Oec Medical Systems, Apr 17, 2002
K022069OEC FLUOROTRAK 9800 PLUS
Ge Oec Medical Systems, Sep 17, 2002
K024012OEC 9800 E/CV+ DIGITAL MOBILE IMAGING SYSTEM
Ge Oec Medical Systems, Mar 20, 2003