Recall Z-0558-2007 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

GE OEC Medical Systems, Inc

OEC 9800 Plus Fluoroscopic X-ray System,

DateMar 22, 2007 (initiated Feb 9, 2007)
StatusTerminated
ReasonDuring routine service, a cover may have been installed without the required proper lead shielding.
Root causeOther
ActionConsignees were notified by telephone on 02/09/2007 and told to discontinue use of the units until they could be inspected by GE personnel. GE personnel visited all 8 sites on 02/12/2007.
Quantity11 units of all models
DistributionNationwide to AL, AZ, IA, OH, TX, VA, and WV.