Recall Z-1105-2007 ยท DWQ Stripper, Vein, External

Smith & Nephew, Inc Endoscopy Division

Smith & Nephew TRIVEX System Pneumatic Footswitch, Part #7209791; distributed by Smith & Nephew, Inc. - Endoscopy, Andover, MA 01810.

DateAug 2, 2007 (initiated Oct 12, 2004)
StatusTerminated
ReasonBuild-up of static charge in Pneumatic Footswitch tubing may cause control unit to reset to default settings in the operating room during surgery.
Root causeOther
ActionFirm sent notification letters to consignees via overnight FedEx on 10/12-13/04. International customers were notified via letter on 11/10/04. Customers were asked to return footswitches for a replacement.
Quantity274 units.
DistributionProduct sent to distributors nationwide and to the following countries: Australia, Austria, Canada, China, Germany, Italy, Japan, and Taiwan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K032387TRIVEX SYSTEM
Smith & Nephew, Inc., Oct 29, 2003