Recall Z-0735-2007 ยท NTI Drug Metabolizing Enzyme Genotyping Systems

Roche Molecular Systems, Inc.

AmpliChip CYP450 Test, CE-IVD, M/N: 4381866190, Roche Molecular Systems, Inc.

DateApr 11, 2007 (initiated Mar 9, 2007)
StatusTerminated
ReasonThe current package insert for the AmpliChip CYP 450 test, lists an incorrect part number for DNase I. The DNase is listed in the "Other Materials Required" section as 'DNase I rec., RNase-free, P/N 04716728001 (Roche Applied Science)" . The Part Number listed for the DNase I listed is incorrect and of lower specific activity than the correct/validated DNase.
Root causeOther
ActionA product advisory notice was issued on March 9, 2007 to all affiliates, indentifying the problem and listing corrective actions followed by a customer letter (Urgent Medical Device Correction) issued on 3/21/2007, requesting several actions be performed including discontinued use of incorrect DNAse I.
Quantity89
DistributionWorld wide Distribution --- USA including states of MI, IN, NC, CA KY, IN, VA, and FL.and countries of Netherlands, Germany, Sweden, Denmark, Italy, Spain, Belgium, Switzerland, and Hungary
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.