Recall Z-0608-2013 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Radiometer America Inc

Qualicheck 1+, Aqueous Quality Control, Level 2, S7140, REF 944-042. The solution is used for quality check of blood gas analyzers.

DateDec 21, 2012 (initiated Jun 6, 2006)
StatusTerminated
ReasonThe barcode for the ABL77 on the insert incorrectly enters a control range for pH of 7.90-7.95 into the analyzer. The insert correctly states that the control range is 7.40-7.45.
Root causeProcess control
ActionRadiometer sent Field Action Notes regarding the affected product. The note identified the product, the problem and the action to be taken by the customer. Instead of scanning the barcorde on the insert to enter the QC data into the anyalyzer, customers were instructed to manually key in the QUALICHECK type, lot number, and correct control range. For questions regarding this recall call 440-871-8900.
Quantity840 Ampules
DistributionDistributed to KY, NC, and NY.