Recall Z-0761-2007 ยท GWQ Full-Montage Standard Electroencephalograph

Bio-Logic Systems Corp

Bio-logic-Ceegraph/Sleepscan Netlink Traveler, Bio-logic Systems Corp., a nebus company, Mundelein, IL 60060; Model 580-T2ASM2. (a Digital EEG/Sleep Recorder Electroencephalograph that includes a built-in pulse oximeter, body position sensor, snore monitor, chest, abdominal and air flow transducers and electrode array for EEG, EMG and EOG)

DateApr 20, 2007 (initiated Feb 26, 2007)
StatusTerminated
ReasonBattery and Battery Compartment; incorrect placement in the Netlink Traveler unit which caused a short in the wiring of the battery pack and overheating.
Root causeOther
ActionOn 2/26/07, Bio-logic Systems shipped replacement batteries, marked with part number 520-LIBAT4-3 and a label reading 'THIS SIDE OUT', and a new battery compartment cover with a label inside the cover to clarify the correct installation of the battery to each of the affected customers. The accounts were requested to remove and replace the old batteries, part 520-LIBTA4-2, and compartment covers, and return the old batteries and covers to Bio-logic Systems in the enclosed pre-paid shipping container, along with the completed accounting form.
Quantity130 battery packs (112 domestic USA, 18 foreign)
DistributionWorldwide, USA , Canada, China, Hong Kong, India and Israel.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K052066CEEGRAPH/SLEEPSCAN NETLINK TRAVELER
Bio-Logic Systems Corp., Aug 26, 2005