Recall Z-0756-2007 ยท MLN Electrode, Electrocardiograph, Multi-Function

Conmed Corporation

PADPRO ADULT Radiotranslucent Pad w/Zoll connector. Defibrillation/Pacing/Cardioversion/Monitoring Electrodes, REF/Catalog # 2516Z. RX ONLY. Made in USA for: CONMED CORPORATION Utica, New York, 13502.

DateApr 24, 2007 (initiated Mar 8, 2007)
StatusTerminated
ReasonFirm was notified of three incidents over the past 15 months in which customers opened the pouch & found an incorrect connector. This error makes the product unusable. The Zoll and Medtronic connectors are not interchangeable and are unique to their specific device.
Root causeOther
ActionOn 3/08/07, "URGENT DEVICE RECALL" letters (dated 3/05/07) will be sent via UPS to the US distributors and direct user accounts. The letters inform them of the problem and of the recall. The letters instruct them to immediately discontinue using the affected lots, inspect their stocks, and to return them as soon as possible. The letters instruct the consignees to notify their customers of the recall of the devices to the user level if they have been distributed, and request the return of the products. Additionally, when requested, ConMed will notify distributor's customers (sub-accounts) by t
Quantity9712 pairs
DistributionWorldwide, including USA, Canada, Romania, Chile, and Belgium
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K020203PADPRO MODEL #2516
Padpro, LLC, May 8, 2002