Recall Z-0718-2007 ยท FRN Pump, Infusion

Smiths Medical MD, Inc.

Deltec branded CADD-Prizm PCS II ambulatory infusion pump and the CADD-Prizm VIP system ambulatory infusion pumps; (Both) Model 6101; Order #21-8861-01 (PCS II) and 21-8821-01 (VIP). Smiths Medical MD, Inc., St. Paul, MN 55112 USA; RX ONLY.

DateApr 5, 2007 (initiated Jan 19, 2007)
StatusTerminated
ReasonThe year of manufacture had been incorrectly entered as 1980 into the pumps software during the manufacture process. As a result the pump may go into an immediate 'Clock battery is Low/Service Immediately' or 'Clock Battery needs service soon' alarm message indicating that the pump requires servicing.
Root causeOther
ActionAll consignees who have received the affected CADD-Prizm PCS II ambulatory infusion pumps and the CADD-Prizm VIP System ambulatory infusion pumps were contacted via telephone by February 1, 2007. The problem will be explained to the consignee and they will be asked to return the pumps to Smiths Medical MD in order for the problem to be corrected.
Quantity27
DistributionNationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K000842CADD-PRIZM PCS II MODEL 6101 AMBULATORY INFUSION SYSTEM
Sims Deltec, Inc., Jun 8, 2000
K960826CADD-PRIZM MODEL 6100 PUMP
Sims Deltec, Inc., Aug 23, 1996
K973917CADD-DIPLOMAT PC COMMUNICATIONS SYSTEM AND THE CADD- AMBASSADOR PUMP COMMUNICATIONS SYSTEM
Sims Deltec, Inc., Jan 9, 1998