Recall Z-0700-2007 ยท LGW Stimulator, Spinal-Cord, Totally Implanted For Pain Relief

Advanced Bionics Corp

The Precision Implantable Pulse Generator, Model Number SC-1110

DateMar 31, 2007 (initiated Jan 30, 2007)
StatusTerminated
ReasonThe Precision Implantable Pulse Generator with firmware Revision 3.01 incorrectly reports an error condition for a valid output from its self internal check.
Root causeOther
ActionMarch 2007 by letter to physicians who had implanted the units notifying them of the recall and future upgrade by wireless RF telemetry.
Quantity5 unimplanted devices; 52 implanted devices
DistributionNationwide including Puerto Rico
FDA recordOpen on fda.gov

510(k) clearances named in this recall