Recall Z-0891-2007 ยท DSI Detector And Alarm, Arrhythmia

General Electric Medical Systems Information Technology

GE Healthcare CIC Pro Clinical Information Center, Central Station, Software Version 4.1. , GE Medical Systems Information Technologies, Inc., 8200 W. Tower Ave. Milwaukee, Wisconsin USA. (System hardware consists of a processing unit and a monitor)

DateJun 6, 2007 (initiated Jan 24, 2007)
StatusTerminated
ReasonAlarm and Reboot problems: Two separate issues could occur with the CIC Pro system when used with Telemetry. Each issue occurs in software version v4.1 only; A) the system may experience a loss of audible alarms when operating system memory resources reach 170MB and/or B) the system may intentionally or unexpectedly reboot .
Root causeOther
ActionGE Healthcare sent a certified letter titled "Urgent Medical Device Correction" to customers on 3/13/07 that explained the short and long term solutions and that the company would be providing a software update that will rectify the issue specific to CIC Pro v4.1 software.
Quantity5,190 Units
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K032370CIC PRO CLINICAL INFORMATION CENTER CENTRAL STATION
Ge Medical Systems Information Technologies, Aug 13, 2003