ClearanceBrief
Recall Z-0785-2007 ยท JCW Prosthesis, Penis, Inflatable

American Medical Systems

AMS 700 Ultrex Preconnected penile Prosthesis with InhibiZone, Product length 18 cm, product number 72403924. Penoscrotal Approach - with InhibiZone, 8 Extenders, one pair each .5cm, 1cm, 2cm 3cm; 1 Tactile Pump, 2 Cylinders.

DateMay 9, 2007 (initiated Feb 9, 2007)
StatusTerminated
ReasonThe carton labeling does not match the product packaged inside.
Root causeOther
ActionAMS has contacted every account (via e-mail 2/9/2007) that either purchased or implanted the affected product. Unused pieces have been recovered and scrapped.
Quantity4
DistributionWorldwide, including USA, Canada and England.
FDA recordOpen on fda.gov

510(k) clearances named in this recall