Recall Z-0549-2007 ยท JAK System, X-Ray, Tomography, Computed
Siemens Medical Solutions USA, Inc.
Symbia T Series SPECT-CT (Single-Photon Emission Computed Tomography and Computed Tomography) System with a Pinhole Collimator; Siemens Medical Solutions USA, Inc., Molecular Imaging, 2501 N. Barrington Road, Hoffman Estates, IL 60195-5203 USA; part number 08717733
| Date | Feb 23, 2007 (initiated Jan 30, 2007) |
| Status | Terminated |
| Reason | Symbia systems, running on e.soft version 5.5 or earlier, could allow a fully-extended patient pallet to come into contact with the pinhole collimator, resulting in system damage. |
| Root cause | Other |
| Action | Siemens sent Customer Safety Advisory Notice/Urgent Field Correction Recall letters dated 1/30/07 to all users of Symbia systems with a pinhole collimator and e.soft software versions 5.5 or earlier. The accounts were informed of the potential for a fully-extended patient pallet to come into contact with the pinhole collimator, resulting in system damage. The customers were advised that a Siemens Customer Service Representative would be sent to their site to upgrade their system software to version syngeo MI Applications 2006A within 30 days. The upgrade will be done free of charge. |
| Quantity | 11 units |
| Distribution | Worldwide distribution --- including USA states of Alabama, California, Georgia, Indiana, Kentucky, Michigan, Ohio, Pennsylvania, Texas, Virginia and countries of Australia, Bolivia, Canada, China, France, Germany, Hong Kong and Spain. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K041166 | SIEMENS ENHANCED IMAGING SYSTEM Siemens Medical Solutions USA, Inc., May 17, 2004 |