Recall Z-0638-2007 ยท CBK Ventilator, Continuous, Facility Use

Viasys Respiratory Care, Inc.dba Bird Products

VIASYS VELA Ventilator, Catalog Number-16186; Viasys Respiratory Care, Inc.

DateMar 20, 2007 (initiated Jan 31, 2007)
StatusTerminated
ReasonThe graphical user interface may become slow in its response and in some cases non-responsive. Under some circumstances this situation may cause the ventilator to delay the annunciation of alarms.
Root causeOther
ActionOn January 31, 2007, a product notification letter was mailed to all customers in the United States via overnight carrier (Federal Express). Consignees are informed in the notification letter that service calls are in process of being scheduled to update domestic consignees. On January 31, 2007, a product notification kit (including a product notification letter, download software, download instruction sheets and installation verification fax-back sheets) were mailed to all international customers via register mail. Consignees are informed in the notification letter that replacement software,
Quantity878
DistributionWorldwide Distribution ---USA and countries of Bangladesh, Bosnia, Brazil, China, Czech Republic, Ecuador, Estonia, Finland, France, Guatemala, Hong Kong, India, Israel, Italy, Japan, Latvia, Lithuania, Malaysia, Morocco, Netherlands, Portugal, Russian Federation, Taiwan, Thailand, Turkey & Vietnam.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K032451BIRD TBIRD VELA VENTILATOR
Bird Products Corp., Mar 17, 2004