ClearanceBrief
Recall Z-0626-2007 ยท IRT Pad, Heating, Powered

HoMedics, Inc.

HoMedics Thera P Model King size Moist/Dry Heating Pad with Auto Shut-off; Model No. HP-500 (UPC 031262003652).

DateMar 21, 2007 (initiated Feb 9, 2007)
StatusTerminated
ReasonHeating pads may have a loose electrical connection, which could result in excessive heat being generated that poses a risk of patient burns, fire, and property damage.
Root causeOther
ActionPress release issued on 2/9/07 instructing consumers to return the product to store of purchase or Homedics. Retail stores notified via recall letter issued on 2/9/07. Retail stores are being instructed to accept returns and to post placards to notify customers of the recall.
Quantity292,108 of all models
DistributionNationwide and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.