ClearanceBrief
Recall Z-0930-2007 ยท DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass

Terumo Cardiovascular Systems Corp

Terumo Advanced Perfusion System 1; 220V -240V, 7A (circuit breaker), 50/60 Hz (10A power source required) base; Model 801764. (Not distributed within the United States).

DateJun 30, 2007 (initiated Feb 1, 2005)
StatusTerminated
ReasonSystem 1 power supply may electrically short to the main console, disrupting normal system operation by causing unexpected re-setting or shutdown of pumps and modules, false alarms and/or overheating and failure of the central control monitor.
Root causeOther
ActionThe firm corrected the complaint units at the time of the complaints in 2005. The firm issued a Device Correction Letter dated 6/6/07 notifying all users of the problem and advising them that their system will be corrected when replacement components become available.
Quantity416 for both models
DistributionWorldwide: USA, Australia, Belgium, Canada, Dominican Republic, Egypt, Germany, Hong Kong, India, Indonesia, Japan, Malaysia, Mexico, Pakistan and Singapore.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022947ADVANCED PERFUSION SYSTEM 1 (100/120 VOLT),ADVANCED PERFUSION SYSTEM 1 (220/240 VOLT, ROLLER PUMPS (6, 4 INCH
Terumo Cardiovascular Systems Corp., Mar 20, 2003