Recall Z-0548-2007 ยท KPS System, Tomography, Computed, Emission

Siemens Medical Solutions USA, Inc.

Symbia S Series SPECT (Single-Photon Emission Computed Tomography) System with a Pinhole Collimator; Siemens Medical Solutions USA, Inc., Molecular Imaging, 2501 N. Barrington Road, Hoffman Estates, IL 60195-5203 USA; part number 08717741

DateFeb 23, 2007 (initiated Jan 30, 2007)
StatusTerminated
ReasonSymbia systems, running on e.soft version 5.5 or earlier, could allow a fully-extended patient pallet to come into contact with the pinhole collimator, resulting in system damage.
Root causeOther
ActionSiemens sent Customer Safety Advisory Notice/Urgent Field Correction Recall letters dated 1/30/07 to all users of Symbia systems with a pinhole collimator and e.soft software versions 5.5 or earlier. The accounts were informed of the potential for a fully-extended patient pallet to come into contact with the pinhole collimator, resulting in system damage. The customers were advised that a Siemens Customer Service Representative would be sent to their site to upgrade their system software to version syngeo MI Applications 2006A within 30 days. The upgrade will be done free of charge.
Quantity16 units
DistributionWorldwide distribution --- including USA states of Alabama, California, Georgia, Indiana, Kentucky, Michigan, Ohio, Pennsylvania, Texas, Virginia and countries of Australia, Bolivia, Canada, China, France, Germany, Hong Kong and Spain.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041166SIEMENS ENHANCED IMAGING SYSTEM
Siemens Medical Solutions USA, Inc., May 17, 2004