ClearanceBrief
Recall Z-0499-2007 ยท J

EMD Chemicals Inc

Anaerotest;Microbiology , 50 test strips 1.15112 Store dry and tightly closed. Store at + 15 C to +25C Merck IVD 64271 Darmstadt, Germany

DateFeb 17, 2007 (initiated Aug 17, 2006)
StatusTerminated
ReasonLot # OC553145 of Anaerotest failed to remain in specification. QC indicator shows an anaerobic environment even if oxygen is present
Root causeOther
ActionA Recall Letter was issued by EMD Chemicals on 8/17/2006 to discontinue product usage and for product return.
Quantity800 packages of 50 strips each.
DistributionProduct was only sent to one US customer: Becton Dickinson, Cayey. Pureto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall