Recall Z-0674-2007 ยท BTI Compressor, Air, Portable
Draeger Medical, Inc.
Medical Air Compressor Model #8413419
| Date | Apr 5, 2007 (initiated Jan 26, 2007) |
| Status | Terminated |
| Reason | Compressors not providing sufficient supply pressure to connected medical ventilator |
| Root cause | Other |
| Action | The recalling firm issued a Medical Device Recall Letter dated 1/22/07 to their customers via Certified Mail. An authorized service representative will inspect each potentially affected device. If a device is found to have an affected dryer assembly, it will be repaired free of charge. |
| Quantity | 67 units |
| Distribution | Nationwide. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K982789 | MEDICAL AIR COMPRESSOR, MODEL # 8413419 Drager, Inc., Oct 29, 1998 |