Recall Z-0814-2007 ยท MDB System, Blood Culturing
bioMerieux, Inc.
BacT/VIEW Software, all versions. in vitro diagnostic database software.
| Date | May 18, 2007 (initiated Dec 19, 2006) |
| Status | Terminated |
| Reason | BacT/VIEW will be unable to run scheduled macro applications unless the user touches the BacT/VIEW touch screen or clicks a function after resynchronizing the clocks. |
| Root cause | Other |
| Action | A customer notification was sent on/about December 19, 2006. Customers will be sent a new BacT/VIEW software version C.30b QTR 1 2007. |
| Quantity | 1233 units |
| Distribution | Worldwide, including USA, Canada, UK, Italy, France, Spain, Germany, Poland, Chile, Japan and the Netherlands. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.