Recall Z-0372-2007 ยท CBK Ventilator, Continuous, Facility Use

MAQUET Inc.

Ventilator System SERVO-i V3.1 PC1777 Maquet Critical Care

DateJan 25, 2007 (initiated May 23, 2006)
StatusTerminated
ReasonReported failures of the SERVO-i ventilators user interface system causing technical errors and increased fault frequencies. Switches may be activated unintentionally, or a function not activate when the switch is selected. Direct access knobs may not function or change the parameter in the wrong direction compared to how they are turned.
Root causeOther
ActionMaquet sent an ''Extended Warranty'' letter on May 23, 2006. On March 13, 2007, Maquet issued a Recall Notification letter to all customers stating that they will replace all PC1777 boards in the serial number range 2400-3300, which have not previously been replaced. A safety alert letter was included outlining the failure indication and effect for the user of the failures, as well as, how to identify whether the Servo-i unit has shown any failure indications.
Quantity1, 970 units
DistributionWorldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K010925SERVOI VENTILATOR SYSTEM, MODEL 64 87 800 E407E