Recall Z-0583-2007 ยท DWE Tubing, Pump, Cardiopulmonary Bypass

Medtronic Perfusion Systems

Medtronic 2T10R1 Intersept Custom Tubing Pack. Pack 2T10R1 consists of two PVC lines; one of them is 132 inch in length and is capped. The second line is 102 inch in length and has a one way vacuum relief valve assembled in one end. Tubing is used in the vent roller heads of the heart lung machine. This tubing is connected to the cannula in the patient in order to divert blood to the circuit

DateMar 8, 2007 (initiated Jan 2, 2007)
StatusTerminated
ReasonOne lot of Custom Perfusion Sets, Catalog 2T10R1 was incorrectly assembled. The one way vacuum relief valve on the 102 inch line is attached backwards, thus preventing flow.
Root causeOther
ActionThere was only one customer. The customer was sent a letter on December 29, 2006, requesting that the impacted product be disposed of or returned to Medtronic for disposal.
Quantity17
DistributionIllinois
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K800178EXTRACORP.'S INTERSEPT CUS. TUBING PK.
Extracorporeal Medical Specialities, Inc., Feb 21, 1980
K891687EXTRACORPOREAL CIRCUIT WITH BIO-ACTIVE SURFACE
Medtronic Blood Systems, Inc., May 31, 1989