Recall Z-0417-2007 ยท DDC Thyroglobulin, Antigen, Antiserum, Control

Beckman Coulter Inc

Access Immunoassay Systems Thyroglobulin Antibody Reagent Kit, Part Number 33890

DateFeb 7, 2007 (initiated Oct 26, 2006)
StatusTerminated
ReasonLow end imprecision affecting both S0 calibrators and patient samples which can lead to inaccurate low level patient results when using the Access Thyroglobulin Antibody assay.
Root causeOther
ActionA Product Corrective Action (PCA) letter was mailed on Oct 27, 2006 to all Thyroglobulin Antibody customers. They are informed that there has been an increase in customer feedback regarding low end imprecision affecting both S0 calibrators and patient samples, which can lead to inaccurate low level patient results when using the Access Thyroglobulin Antibody assay. The letter informs the customer that the firm recommends that they use an alternate method for testing TgAb. (Letter was sent by US mail). Customers are asked to complete and return the enclosed response form within 10 days.
Quantity474 units
DistributionNationwide and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K012208ACCESS THYROGLOBULIN ANTIBODY AND CALIBRATORS ON THE ACCESS IMMUNOASSAY SYSTEMS, MODELS 33890, 33895
Beckman Coulter, Inc., Sep 7, 2001