ClearanceBrief
Recall Z-0803-2007 ยท JAK System, X-Ray, Tomography, Computed

GE Healthcare

GE LightSpeed VCT Select / LightSpeed Pro32, Model 5126093, CT Scanner system.

DateMay 15, 2007 (initiated Dec 21, 2006)
StatusTerminated
ReasonOil may leak onto the port window of the CT tube and may result in image artifacts that could lead to misdiagnosis.
Root causeOther
ActionA Field Modification Instruction was issued to the field to inspect all systems and replace the tube window screws and the window filter, as well as clean any traces of oil in the tube window area. A 12/21/2006, Urgent Product Correction Notice/Customer letter has also been distributed informing the customers of the safety issue and the product correction.
Quantity158
DistributionWorldwide Distribution: USA in the following states and locations: AL, AK, AZ, AR, CA, CO, CT, DE, DC, FL, GA, HI, ID, IL, IN, IA, KS, KY, LA, ME, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NY, NC, ND, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WV, WI, Guam and Puerto Rico. Foreign c
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K040372LIGHTSPEED 7.0 CT SCANNER SYSTEM
Ge Medical Systems, LLC, Mar 1, 2004