Recall Z-0600-2007 ยท DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass
Terumo Cardiovascular Systems Corp
Terumo Advanced Perfusion System 1, 4 inch diameter (small) Roller Pump; Model 801040.
| Date | Jun 30, 2007 (initiated Sep 1, 2004) |
| Status | Terminated |
| Reason | The internal ribbon cable may loosen from the roller pump display board, resulting in loss of local control and/or local display. |
| Root cause | Other |
| Action | The firm corrected units in the field or brought them back for evaluation and repair as complaints were received between 9/04 and 4/06. The firm issued a Device Correction Letter dated 6/6/07 notifying all users of the problem and advising them that they will be visited and their units upgraded as replacement components become available. |
| Quantity | 1,484 of both models |
| Distribution | Worldwide: USA, Australia, Belgium, Canada, Dominican Republic, Egypt, Germany, Hong Kong, India, Indonesia, Japan, Korea, Mexico, Malaysia, Singapore, South Africa, Taiwan, Thailand and United Arab Emirates. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K022947 | ADVANCED PERFUSION SYSTEM 1 (100/120 VOLT),ADVANCED PERFUSION SYSTEM 1 (220/240 VOLT, ROLLER PUMPS (6, 4 INCH Terumo Cardiovascular Systems Corp., Mar 20, 2003 |