Recall Z-0383-2007 ยท DQY Catheter, Percutaneous

Boston Scientific

Boston Scientific Mach 1 Guide Catheter Femoral Approach, 6F, 0.070'', Voda Left Curve, UPN # H74934356710, Catalog # 34356-71. Made in Mexico: Edificio G1 Local B, H, Parque Industrial La Mesa, Tijuana, Baja California 22650 MEX.

DateFeb 7, 2007 (initiated Dec 13, 2006)
StatusTerminated
ReasonExcess strands of resin may exist in the lumen near the hub of certain batches/lots of the Mach 1 Guide Catheters. If this excess resin is present in the catheter, there is the potential of an embolization resulting from the excess resin detaching from the device during the procedure.
Root causeOther
ActionAn Urgent Medical Device Recall letter, dated 12/13/06, and a Clarification of Recall Notification letter, dated 12/14/06, were sent to all affected hospitals. The letter describes the issue, affected product, asks for customers to discontinue further distribution or use of any remaining product and asks for product to be returned. Replacement product will be available for any/all returned product.
Quantity9
DistributionNationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K0108746F MACH 1, MODEL 34356-XXX
Boston Scientific Corp, Jun 21, 2001