ClearanceBrief
Recall Z-0485-2007 ยท FMT Warmer, Infant Radiant

Fisher & Paykel Healthcare Ltd

Fisher & Paykel IW930 CosyCot Infant Radiant Warmer (with integrated bed) fitted with 9001W005J Elevator Module

DateJan 11, 2007 (initiated Oct 9, 2006)
StatusTerminated
ReasonThis firm has received Product Complaint Reports concerning bending of the pivot plate. The elevator base of the CosyCot Infant Warmer incorporates a pivot plate which forms part of the mechanism allowing the CosyCot to raise and lower in height. A bent pivot plate generally results in the Warmer being unable to elevate properly. Other component failures may result.
Root causeOther
ActionFisher & Paykel Healthcare commenced sending out Product Correction Notifications letter via U.S. Postal Service certified mail October 9, 2006. This Product Correction Notice explained the nature of the problem and include a reminder of the maximum recommended weight of accessories. Two FAX Response Forms were provided, Part A for initial receipt of the Product Correction Notice and Part B for confirmation of the completed upgrade.
Quantity273 units in US & 9 in Canada
DistributionNationwide and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K970432INFANT RADIANT WARMER - MODELS IW910 AND IW920
Fisher & Paykel Electronics , Ltd., May 7, 1997