ClearanceBrief
Recall Z-0360-2007 ยท FTL Mesh, Surgical, Polymeric

Davol, Inc., Sub. C. R. Bard, Inc.

Bard Composix Kugel Hernia Patch- Large Circle 4.5'' with ePTFE Product Code: 0010204

DateJan 30, 2007 (initiated Jan 10, 2007)
StatusCompleted
ReasonPET recoil ring breakage-Memory recoil ring may break which could potentially lead to bowel perforation and or chronic enteric fistula
Root causeOther
ActionDavol Inc. notified Hospital Administrators/Risk Health Managers,distributors, and surgeons of the expanded recall by letter beginning 1/10/07. Product under recall is to be returned to Davol for replacement. Davol is posting recall information to their website.
Quantity16,603 units
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K003323COMPOSIX KUGEL MESH, MODEL 0010201,0010202,0010203,0010204,0010205
Davol, Inc., Jan 22, 2001