Recall Z-0970-2007 ยท BZD Ventilator, Non-Continuous (Respirator)
Fisher & Paykel Healthcare Inc
Fisher & Paykel Aclaim Nasal Mask, Part Numbers: HC401A & 900HC401. CPAP masks.
| Date | Jun 27, 2007 (initiated Nov 24, 2006) |
| Status | Terminated |
| Reason | Plastic tabs incorporated into the connector components of these CPAP masks may break off if they are not cleaned in accordance with the Instructions for Use. In the event of a tab breaking off, it may enter the CPAP system air path and there is an extremely remote possibility that the broken tab may then enter the patient's nasal cavity or mouth. Masks manufactured since April 2006 do not feat |
| Root cause | Other |
| Action | Fisher & Paykel Healthcare Inc. commenced sending out letters by mail to consignees the week commencing 4 December 2006 advising consignees to quarantine and destroy affected product. A faxback form was also be provided for consignees to advise Fisher & Paykel Healthcare of quarantined and destroyed stock. The firm will supply consignees with replacement product as determined from completed faxback information. The firm will collate responses from consignees and will follow up with a letter to consignees who do not respond within a month of initial mail-out. The firm will then make telephone |
| Quantity | 815,239 total for all models, all lots |
| Distribution | Nationwide |
| FDA record | Open on fda.gov |