Recall Z-0356-2007 ยท CFN Method, Nephelometric, Immunoglobulins (G, A, M)

Roche Diagnostics Corp.

Roche IGGT COBAS INTEGRA Immunoglobulin G (Turbidimetric); Catalog no. 20766631322. in vitro diagnostic.

DateJan 17, 2007 (initiated Dec 15, 2006)
StatusTerminated
ReasonFalsely elevated IGGT results may be reported because the R2 reagent of the Tina-quant Gen.2 albumin assay has a carryover effect on the Immunoglobulin G (turbidimetric) assay on the COBAS INTEGRA 700 and 800 analyzers and the R1 reagent of the Tina-quant Hemoglobin A1c assay has a carryover effect on the Immunoglobulin G (turbidimetric) assay on the COBAS INTEGRA 800 analyzer.
Root causeOther
ActionDevice correction letters dated 12/15/06 instructing users to perform an extra wash cycle between applications when using both the ALBT2 and IGGT cassettes on the COBAS INTEGRA 700 and 800 analyzers. The recall was extended to via device correction letters dated 2/28/07 instructing users to perform an extra wash cycle between applications when using both the A1C2 and IGGT cassettes on the COBAS INTEGRA 800 analyzer.
Quantity18,905
DistributionNationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K982382COBAS INTEGRA REAGENT CASSETTES
Roche Diagnostic Systems, Inc., Sep 28, 1998