ClearanceBrief
Recall Z-0633-2007 ยท FRN Pump, Infusion

Baxter Healthcare Corp.

Baxter Flo-Gard 6301 Dual Channel Volumetric Infusion Pump, catalog #2M8064, 2M8064R (refurbished); Baxter Healthcare Corporation, Deerfield, IL 60015

DateMar 20, 2007 (initiated Nov 27, 2006)
StatusTerminated
ReasonThere is the potential for non-detection of, and no alarm for, upstream occlusion for Flo-Gard Volumetric Infusion Pumps, if an occlusion occurs above a flexible chamber such as a drip chamber, Buretrol or blood filter.
Root causeOther
ActionBaxter sent the 11/27/06 Urgent Device Correction letter to all Flo-Gard Volumetric Infusion Pump customers, to the attention of the Vice President of Nursing and the Biomedical Engineer, via first class mail to alert them to the potential for non-detection of, and no alarm for, upstream occlusion under specific conditions for Flo-Gard Volumetric Infusion Pumps if the occlusion occurs above a flexible chamber (such as a drip chamber, Buretrol or blood filter), and to provide the addendum to the operator manual to address these specific conditions. The letters included the mitigations and risk
Quantity42,098 pumps
DistributionWorldwide including USA, Puerto Rico, Argentina, Australia, Bahrain, Belgium, Bermuda, Brazil, Canada, Chile, China, Colombia, Costa Rica, Denmark, Ecuador, Egypt, El Salvador, Guatemala, Hong Kong, Honduras, India, Indonesia, Ireland, Italy, Korea, Lebanon, Malaysia, Mexico, Netherlands, New Zealan
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K915523FLO-GARD(R) 6301
Baxter Healthcare Corp, Feb 28, 1992