Recall Z-0504-2007 ยท IXR Table, Radiographic, Tilting

Stille-Sonesta, Inc

Stille Sonesta Video Fluoroscopy Table 6210; manufactured by Stille AB, Solna, Sweden.

DateFeb 15, 2007 (initiated Nov 9, 2006)
StatusTerminated
ReasonBolts may back out of their joints causing the back section of the table to tilt.
Root causeOther
ActionNotifications letters sent to consignees on 11/20/06. Users are instructed to check the bolts located under the table. The Notice includes instructions with pictures for checking the bolts and tightening them if necessary. Stille AB has developed a kit for replacing the bolt that is involved and will send a service engineer out to the site to do the replacement.
Quantity66 units.
DistributionNationwide.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.