Recall Z-0454-2007 ยท DLZ Enzyme Immunoassay, Phenobarbital

Abbott Laboratories Inc.

Phenobarbital- Reagent-Clinical Chemistry--List number: 1E08-20-Abbott Diagnostics Division

DateJan 9, 2007 (initiated Sep 26, 2006)
StatusTerminated
ReasonErratic elevated results and or the inability to calibrate due to imprecision.
Root causeOther
ActionAbbott sent out a product recall letter on 9/26, 2006, to all consignees that received lot numbers 39074HW00, 41008HW00 and 42063HW00. The letter states that the firm has identified a Phenobarbital performance issue that is demonstrated by erratic elevated results and/or the inability to calibrate due to imprecision. Until the cause of this issue is identified, they have discontinued shipment of Phenobarbital reagent. It will be necessary to identity an alternate means for Phenobarbital testing during this time. AXSYM and TDxITDxFLx Phenobarbital are available as alternate testing options (co
QuantityWorldwide: Lot #41008HW00 - 162 kits; Lot #42063HW00 - 136 kits; Lot #39074HW00 - 150 Kits
DistributionWorldwide distribution ---- USA and countries of of Canada, Chile, Germany, Australia
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K993031ABBOTT AEROSET PHENOBARBITAL ASSEY AND CALIBRATORS;MODELS 1E08 AND 5F54
Syva Co., Jan 27, 2000