Recall Z-0288-2007 ยท CXT

Mallinckrodt Inc

OptiVantage DH Power Injection System.

DateDec 16, 2006 (initiated Nov 7, 2006)
StatusTerminated
ReasonThe flow rate programmed on the console may be changed without automatically updating the flow rate programmed on the power head. The injector will always inject in accordance with the parameters shown on the power head. If this lack of synchronization occurs and is not noticed by the clinican, contrast would be injected at a higher or lower flow rate than desired.
Root causeOther
ActionThe recalling firm sent a letter dated 11/7/2006 to all consignees.
Quantity1348 devices.
DistributionWorldwide distribution ---- including USA and countries of Australia, Belgium, Brazil, Canada, Chile, Columbia, Denmark, Mexico, Panama, Puerto Rico, Singapore and South Africa.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K042744OPTIV ANTAGE DH POWER INJECTION SYSTEM
Mallinckrodt Inc., Liebel-Flarsheim Business, Nov 30, 2004