Recall Z-0481-2007 ยท CBK Ventilator, Continuous, Facility Use

Viasys Respiratory Care, Inc.

Viasys V.I.P. Bird Sterling pediatric ventilator, Catalog Number 15654

DateJan 11, 2007 (initiated Oct 20, 2006)
StatusTerminated
ReasonThe reason for this correction is due to the potential of a repeat ''failure to cycle'' event.
Root causeOther
ActionOn November 2, 2006, a product notification letter was mailed to all customers in the United States via overnight carrier (Federal Express). Consignees are informed in the notification letter that service calls are in process of being scheduled to update domestic consignees. On November 4, 2006, a product notification letter was mailed to all international customers via registed mail. Consignees are informed in the notification letter that replacement software, replacement instructions, and installation verification forms are included in their recall package.
Quantity539
DistributionWorldwide including USA, Canada, Estonia, Belgium, Saudi Arabia, Brazil, Thailand, United Arab Emirates, Syria, Portugal, Jordan, Italy, Latvia, Hong Kong China, Ecuador, Jamaica, Dominican Republic, New Zealand, Australia, Czech Republic, Bolivia, Austria, Greece, United Kingdom (England), Turkey
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K993449BIRD V.I.P. GOLD/STERLING
Bird Products Corp., Dec 20, 1999