ClearanceBrief
Recall Z-0422-2007 ยท IYN System, Imaging, Pulsed Doppler, Ultrasonic

Philips Ultrasound, Inc.

iE33 Ultrasound System (System, imaging, pulsed doppler, ultrasonic and system, imaging, pulsed echo, ultrasonic)

DateFeb 17, 2007 (initiated Nov 2, 2006)
StatusTerminated
ReasonAcoustic output intensity may exceed maximum specified limits and the patient contact temperature may exceed 43 degree C (when the system is used with the S12-4 transducer in CW-mode)
Root causeOther
ActionOn 11/2/06 the firm sent a letter dated October 31, 2006 to their customers. The letter explains the problem and states that new software is being released and will be available sometime in November. Customer service representatives will visit consignees to upgrade their units. The firm later determined they sent letters out to consignees that did not have affected units. Those consignees will be notified the week of 11/20/06 to disregard the previous letter.
Quantity619 domestic units, 706 international units
DistributionWorldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K030455BORIS PLATFORM DIAGNOSTIC ULTRASOUND SYSTEM
Philips Ultrasound, Inc., Mar 13, 2003