Recall Z-0449-2007 ยท GKZ Counter, Differential Cell

Drew Scientific Group PLC

Drew Excell 22 Hematology Analyzer, manufactured by Drew Scientific, Inc., Dallas, TX.

DateJan 9, 2007 (initiated Aug 24, 2006)
StatusTerminated
ReasonFlags and alerts on the user interface will not automatically transfering to the LIS.
Root causeOther
ActionFirm sent 'Product Notification #22' to all customers on 08/24/06. Users advised of potential risk to patient health and provided with instructions on how to eliminate risk by disabling the auto transfer feature and manually transfering data until a software patch is made available
Quantity299 units.
DistributionWorldwide distribution --- including USA states of FL, MA, MN, ND, NM, NY, Puerto Rico and TX and countries of Austria, Bangladesh, Chile, China, Czech Republic, Egypt, France, Hong Kong, Italy, India, Iran, Korea, Mexico, Morocco, Phillipines, Portugal, Russia, Serbia and Montenegro, Spain, Syria,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K991539EXCELL 22
Mwi, Inc., Jul 1, 1999