Recall Z-0341-2007 ยท LYJ Stimulator, Autonomic Nerve, Implanted For Epilepsy

Cyberonics, Inc

Cyberonics VNS Therapy System, Model 250-''HAND HELD'' programming software v6.1, v6.1.7, v7.0 and v7.1.3; manufactured by Cyberonics, Inc., Houston, TX.

DateJan 25, 2007 (initiated Oct 17, 2006)
StatusTerminated
ReasonDuring programming, pulse generator may be inadvertently set to 8.0 mA output, regardless of the mA range selected by the clinician
Root causeOther
ActionFirm sent safety alert to physicians on 11/14/06. Safety Alert informs physicians of the problem and provides instructions on how to avoid an event and correct an event should it occur.
Quantity3,538 Flashcards containing software.
DistributionWorldwide Distribution ---- USA and countries of Argentina, Australia, Austria, Belgium, Canada, Cyprus, Czech Rep, Denmark, Finland, France, Germany, Greece, Hong Kong, Hungary, Iceland, Ireland, Israel, Italy, Japan, Jordan, Malaysia, Mexico, Netherlands, Norway, Poland, Portugal, S. Korea, Sloven