Recall Z-0221-2007 ยท ITX Transducer, Ultrasonic, Diagnostic

Computerized Medical Systems Inc

AccuSeed DS Digital Stepper, an accessory to a prostate radiation treatment planning system. Model #DS300

DateNov 29, 2006 (initiated Nov 9, 2006)
StatusTerminated
ReasonThe In/Out adjustment knob can be over-turned, ultimately causing the incorrect placement of radioactive seeds in the patient.
Root causeOther
ActionThe recalling firm issued letters dated 11/9/06 via regular mail. The letter informs the customer of the problem and includes an advisory notice instructing them to perform a test prior to each use. If the system does not function correctly, it can still be used safely in a manual mode until it can be returned for repair. A questionnaire was enclosed that will facilitate the firm in prioritizing repairs and retrofits. After the questionnaire is reviewed, the firm will contact the customer for further information regarding their facilities use of the device. This letter was also issued
Quantity62 steppers
DistributionWorldwide, including USA, Australia, Japan, China, France, Germany and Brazil.